-40%
Multifunction Thermocouple RTD Process Calibrator 2in1 DCV Ohm Source Output
$ 250.8
- Description
- Size Guide
Description
Note: USB function is not available.Multifunction Temperature
Calibrator
(referred to as
“The Calibrator”
)
is a handheld, high capacity and rechargeable battery-operated instrument that measures and sources electrical and physical parameters. Such as measures and simulates a variety of thermocouple (TC) and Resistance Temperature Detector
(RTD).
See Table 1.
In addition to the functions listed in Table 1, the Calibrator also has the following characteristics
1
.
A split-screen display. The upper display allows users to measure volts, current. The lower display allows the user to measure and source Millivolt, resistance temperature detectors, thermocouples and ohms.
2
.
A thermocouple (TC) input/output terminal with Cold Junction Compensation, ambient temperature can be set and calibrate within the menu function.
3
.
Each parameters can manual and automatic stepping and ramping output, signal output can be arbitrarily set within the 0% and 100% of the value.
4
.
In the configuration menu can be set manually: Power off time for Instrument, Backlight off time, Screen Contrast, Cold Junction compensation ON/OFF, HART resistance ON/OFF and etc.
5
.
Have calendar clock function, it can be calibrated in configuration menu.
6. It can be store 8 sets of measurement and output data and call the showing at any time, all the settings and the functions of the last operation be auto-stored after power off.
7. High capacity rechargeable battery pack with power adapter within integrated charge-management functions.
8
.
Novel shape and large screen dot-matrix graphic display.
Table 1. Summary of Source and Measure Functions
Function
Measure signal
Output
DC Voltage (the upper display)
0.000-30.000 V
DC Current (the upper display)
0.000-24.000 mA
Loop
power (the upper display)
0.000-24.000 mA
DC Millivolt
-10.000mV-80.000mV
-10.000mV-80.000mV
Resistance
0Ω to 400.00Ω
400.0Ω to 4000.0Ω
5.00Ω to 400.00Ω
400.0Ω to 4000.0Ω
Thermocouple (TC)
Type: S, R, B, K, N, E, J, T(mV)
Resistance Temperature Detector (RTD)
Pt100(385) Pt100(3926) Pt100(3916) Pt200(385)
Pt500(385) Pt1000(385) Cu100(ITS) Cu50(ITS)
Other functions
Stepping output, Ramping output, Store data, a split-screen display.
Technical Index:
Specifications are based on a one year calibration cycle and apply from +
18 °C
to +
28 °C
unless stated otherwise. All specifications assume a five-minute warmup period.
The presentation of accuracy is: ± (reading value *% + lower order digits)
Temperature coefficient:
-10 ℃
to
18 ℃
, +
28 ℃
to
55 ℃
: ±0.005 % of range per °C
1
.
DC Voltage Measurement
Range
Resolution
Accuracy
0-30.000V
(
upper display
)
0.001V
0.02%+3
-10.000-80.000mV
0.001mV
0.01%+10
2
.
DC Voltage output
Range
Resolution
Accuracy
-10.000—80.000mV
0.001mV
0.01%+10
3
.
DC Current (Milliampere) Measurement
Range
resolution
Accuracy
0—24.000mA
(
upper dispaly
)
0.001mA
0.02%+4
4.
Ohms Measurement
Ohms
Range
Accuracy±Ω
4-wire
(
4W
)
2-wire
(
2W
)
and 3-wire
(
3W
)
0—400.00Ω
0.1Ω
0.15Ω
400.0—4000.0Ω
0.5Ω
1Ω
Excitation Current: 0.2mA.Maximum input voltage: 30V.2-wire: Does not include lead resistance.
3-wire:
Assumes matched leads with a total resistance not exceeding 100 Ω.
5. Ohms Source
Range
Excitation Current from Measurement Device
Accuracy±Ω
Resolution
5.00Ω to 400.00Ω
0.15mA-0.5mA
0.1Ω
0.01Ω
5.00Ω to 400.00Ω
0.5mA-2mA
0.05Ω
0.01Ω
400.0 to 1500.0Ω
0.05mA-0.8mA
0.8Ω
0.1Ω
1500.0Ω to 4000.0Ω
0.05mA-0.4mA
1.5Ω
0.1Ω
6.
Temperature ,Thermocouples
Type
Range
Resolution
Measurement and Source Accuracy±
℃
S
-50.0
℃
to
0
℃
0.1
℃
/
0.1
F
1.5
℃
0
℃
to
500.0
℃
0.1
℃
/
0.1
F
1
℃
500.0
℃
to
1760.0
℃
0.1
℃
/
0.1
F
0.8
℃
R
-50.0℃
to
0℃
0.1℃
/
0.1℉
1.5℃
0℃
to
500.0℃
0.1℃
/
0.1℉
1℃
500.0℃
to
1760.0℃
0.1℃
/
0.1℉
0.8℃
B
200℃
to
800℃
800℃
to
1800℃
1℃
/
1℉
2.2℃
1.8℃
K
-200.0℃
to
1370.0℃
0.1℃
/
0.1℉
0.8℃
N
-200.0℃
to
1300.0℃
0.1℃
/
0.1℉
0.8℃
E
-200.0
℃
to
1000.0℃
0.1℃
/
0.1℉
0.5℃
J
-200.0℃
to
1200.0℃
0.1℃
/
0.1℉
0.5℃
T
-200.0℃
to
400.0℃
0.1℃
/
0.1℉
0.5℃
The thermocouple complies with ITS-90.
If open the cold junction compensation (
CJC
ON
), additional plus ±
0.3 ℃
.
7.Temperature, Resistance Temperature Detectors (RTD)
Type
Range
Accuracy±
℃
Measure 4-wire
℃
Measure 2-wire and 3-wire
℃
Output
℃
Pt100-385
-200.0
℃
-850.0
℃
0.3
℃
0.5
℃
0.3
℃
Pt100-3926
-200.0
℃
-850.0
℃
0.3
℃
0.5
℃
0.3
℃
Pt100-JIS
-200.0
℃
-850.0
℃
0.3
℃
0.5
℃
0.3
℃
Pt200-385
-200.0
℃
-250.0
℃
0.2
℃
0.3
℃
0.2
℃
250.0
℃
-630.0
℃
0.8
℃
1.6
℃
0.8
℃
Pt500-385
-200.0
℃
-500.0
℃
0.3
℃
0.6
℃
0.3
℃
500.0
℃
—
630.0
℃
0.4
℃
0.9
℃
0.4
℃
Pt1000-385
-200.0
℃
-100.0
℃
0.2
℃
0.2
℃
0.3
℃
100.0
℃
-630.0
℃
0.2
℃
0.3
℃
0.3
℃
Cu100
-50.0℃
-150.0℃
0.4℃
0.6℃
0.4℃
Cu50
-50.0℃
-150.0℃
0.4℃
0.6℃
0.4℃
Resolution: 0.1 ° C, 0.1 ° F.
Allowable excitation current (Source): Pt100-385, Pt100-392, Pt100-JIS, Pt200-385: 0.15 to 3.0 mA
Pt500-385: 0.05 to 0.80mA; Pt1000-385: 0.05 to 0.40mA.
*
2-wire: Does not include lead resistance.
*3-wire: Assumes matched leads with a total resistance not exceeding 100 Ω.
8. Loop Power
Voltage: 24 V; Maximum current: 22 mA ; Short circuit protected.
9. Complex Index
Operating temperature
-10
℃
to
50
℃
Storage temperature
-20
℃
to
70
℃
Operating altitude
3000 meters above mean sea level
Relative Humidity (% RH operating without condensation)
75%(
30℃
to
40℃
) / 45%(
40℃
to
50℃
)
35%(
50℃
to
55℃
) / uncontrolled <
10 °C
Stability
± 0.005 % of range/°C outside of 23 ± 5 °C
Vibration
Random,
2g
, 5 to 500 Hz
Safety
EN 61010-1:1992
Protection Class
Pollution Degree II
Protection Class
(
L*W*H
)
224mm
*
172mm
*
65mm
Weight
About
1.5kG
(Include the battery)
Accessories:
a pair of Test leads, a power adapter(US, Euro, UK or AU plug), 8PCS x
Ni-MH rechargeable battery 1.2V AA, a t
hermocouple conversion plug, Carrying case. PDF manual.